| Catalog Number |
150600006 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/03/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's right knee was revised due to loosening of both his primary femoral and tibial components at the cement to implant interface.Depuy cement was used.During the revision procedure, the patient's primary attune components were removed and replaced by attune revision components.There was no cement remaining on the explanted components.The tibial tray and femoral component were loosened and debonded at the cement to implant interface.Affected side: right knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: it was reported that the patient's right knee was revised due to loosening of both his primary femoral and tibial components at the cement to implant interface.Depuy cement was used.During the revision procedure, the patient's primary attune components were removed and replaced by attune revision components.There was no cement remaining on the explanted components, so surgeon seems to think that something may have happened during the mixing of the cement that led to the cement not adhering to the cemented components.The product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned found that the attune fb tib base sz 6 cem has signs of burnishing on the bottom surface, which usually suggests micromotion between the cement and the implant, therefore, together with the lack of debris/cement it is reasonable to confirm loosening.With the information provided is not possible to determine a potential cause at this moment.The mode of failure of the device is multi-factorial and consideration must be given to all other potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the attune fb tib base sz 6 cem would have contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|