• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA - PRINCETON DURAGEN PLUS DURAL REGENERATION MATRIX 1X3 5 PAC; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA - PRINCETON DURAGEN PLUS DURAL REGENERATION MATRIX 1X3 5 PAC; N/A Back to Search Results
Catalog Number DP5013
Medical Device Problem Code Failure to Seal (4070)
Health Effect - Clinical Code Cerebrospinal Fluid Leakage (1772)
Date of Event 12/23/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
This is 1 of 3 reports linked to mfg report numbers 1121308-2025-00015 and 1121308-2025-00017: a facility reported that dura leaks (csf leakage) with the same lot number of the duragen plus dural regeneration matrix (dp5013) occurred post-surgery.No, post-operation complication occurred after the procedure (linked to the same lot number), and no duragen has been removed from patients.However, duragen products with the same lot number has been removed from circulation.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields: a2, a3a, a4, b7, d4, d9, d10, g3, g6, h2, h3, h4, h6, h11.Additional information has been received for incident occurring on 12/23/2024: what revision/medical intervention was performed after csf leak occurred? answer: hospitalization for observation.How much csf leaked during the incident (if an exact amount cannot be provided, please provide an approximate amount: small amount, moderate, etc.)? answer: moderate.Was the patient moved, or being moved at the time of the incident? answer: unknown.Please provide patient outcome answer: expected outcome for which surgery was performed.What type of procedure was duragen used for? answer: suboccipital crani for microvascular decompression.The timeframe of implant date to incident date.Answer: unknown.Details surrounding the intra-operative procedure.Answer: the microscope was removed.The dura was reapproximated with interrupted 4-0 nurolon's.We placed a small duragen overlying this.The craniotomy was then plated with titanium plates and screws including 2 bur hole covers and 3 long dog bones.This completely covered the craniotomy defect.The advaseal was now injected underneath the craniotomy and bur hole covers.The entire wound was now copiously irrigated with lr.We closed in layered fashion.Surgical technique (sterile, documented surgical class as clean procedure).Any concomitant devices (i.E.Surgical sealants) used.Answer: vistaseal.Investigation findings: duragen plus dural regeneration matrix (dp5013) was not returned for evaluation for the units involved in the six (6) cases reported.According to the complaint information no product will be returned and the product with the same lot number was removed from circulation.Also, no photo showing the reported condition and/or product was provided; thus, the reported lot number cannot be confirmed either.However, an investigation of retained samples was performed: device history record (dhr) review - the dhr was reviewed, and no anomalies were found during the manufacturing and packaging process that could be related to the reported condition.Retained sample failure analysis - five (5) units were visually inspected.The dispenser box and contents were in good condition, tray sealing area was complete in each unit, no defects were observed on any of the five (5) sponges, and no foreign material presence was observed.Retained sample medical assessment - a medical assessment was performed to evaluate the possible relation between the reported event and product use and concludes as follows: "there is insufficient information to conclude whether the reported adverse event of cerebrospinal fluid (csf) leak post-surgery was due to the duragen plus.The reported complaints lack detailed patient and event information, including the timeframe of implant date to incident date, patient¿s relevant medical history, details surrounding the intra-operative procedure, surgical technique, and any concomitant devices (i.E.Surgical sealants) used.Further information has been requested by the complaints department but to date, no additional response has been received from the customer." retained sample root cause - based on the available information, there is insufficient information to conclude whether the reported adverse events of cerebrospinal fluid (csf) leak post-surgery were due to the duragen plus.All reports are from a common source which may suggest that the situation may be technique and/or user related.For now, integra will continue to monitor for possible recurrence for trending purposes and to evaluate if any actions are required in the future.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURAGEN PLUS DURAL REGENERATION MATRIX 1X3 5 PAC
Common Device Name
N/A
Manufacturer (Section D)
INTEGRA - PRINCETON
1100 campus road
1100 campus road
princeton NJ 08540
Manufacturer (Section G)
INTEGRA - PRINCETON
1100 campus road
none
princeton NJ 08540
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 08540
6099362319
MDR Report Key21926496
Report Number1121308-2025-00016
Device Sequence Number9030949
Product Code GXQ
UDI-Device Identifier10381780070924
UDI-Public10381780070924
Combination Product (Y/N)N
PMA/510(K) Number
K032693
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberDP5013
Device Lot Number7377759
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/02/2025
Supplement Date Received by Manufacturer05/22/2025
Initial Report FDA Received Date04/29/2025
Supplement Report FDA Received Date06/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VISTASEAL
Outcome Attributed to Adverse Event Other;
Patient Age29 YR
Patient SexFemale
Patient Weight107 KG
-
-