• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. INTERTAN 7.0MM COMP SCREW STARTER DRILL; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. INTERTAN 7.0MM COMP SCREW STARTER DRILL; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 71674570
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/06/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an internal fixation surgery, the tip of one (1) intertan 7.0mm comp screw starter drill broke off while drilling in the lateral cortex.The surgeon removed it from the lateral cortex with a needle driver.It was confirmed that all metal material was removed with x-ray.The procedure was resumed, after a non-significant delay, with a s+n back-up device.No injury was reported as a consequence of this issue.
 
Additional Manufacturer Narrative
H11: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
Corrected data: h6 (component code, investigation findings and investigation conclusions).Additional information: h6 (type of investigation), h11 (roi).H3, h6: the device was not returned for evaluation; therefore.However, the photographs were reviewed and revealed that the tip of the device broke off.The device shows signs of wear and use.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed a similar event for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERTAN 7.0MM COMP SCREW STARTER DRILL
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key21928020
Report Number1020279-2025-00736
Device Sequence Number9086118
Product Code HSB
UDI-Device Identifier00885556389393
UDI-Public885556389393
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71674570
Device Lot Number17BNG0003
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/06/2025
Supplement Date Received by Manufacturer05/21/2025
Initial Report FDA Received Date04/29/2025
Supplement Report FDA Received Date05/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/03/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-