• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERATHON MEDICAL ULC BFLEX 2 REGULAR 5.0 SINGLE-USE BRONCHOSCOPE; BRONCHOSCOPE (FLEXIBLE OR RIGID)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC BFLEX 2 REGULAR 5.0 SINGLE-USE BRONCHOSCOPE; BRONCHOSCOPE (FLEXIBLE OR RIGID) Back to Search Results
Model Number 0570-0423
Medical Device Problem Codes Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer was unable to have the reported bflex 2 5.0 single-use bronchoscope returned to verathon for evaluation as it had been discarded at the facility, therefore the cause could not be determined.Review of complaint history for the reported bronchoscope lot number did not identify any previous complaints reported to verathon.Verathon confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized per the system risk assessment.Trending analysis for the bflex 2 5.0 single-use bronchoscopes does not identify any trends exceeding acceptable limits.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event or Problem Description
A customer reported that during a patient procedure, using a bflex 2 regular 5.0 single-use bronchoscope, the image on the connected glidescope core monitor was flickering in and out.They tried using the bronchoscope on a different glidescope core monitor; however, the issue persisted.No delay in the procedure, use of a backup device, or harm to the patient was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BFLEX 2 REGULAR 5.0 SINGLE-USE BRONCHOSCOPE
Common Device Name
BRONCHOSCOPE (FLEXIBLE OR RIGID)
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key21928739
Report Number9615393-2025-00043
Device Sequence Number9031017
Product Code EOQ
UDI-Device Identifier00879123008565
UDI-Public(01)00879123008565(10)KS244102(11)241009(17)271011
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K183256
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0570-0423
Device Catalogue Number0570-0449
Device Lot NumberKS244102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2025
Initial Report FDA Received Date04/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-