It was reported that the procedure performed was to treat mildly calcified, mildly tortuous left anterior descending coronary artery.Once at the lesion, the 4.00x15mm trek balloon ruptured at 11 atmospheres during the 2nd inflation.A non-abbott device was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.Based on the information received, the investigation determined that the reported balloon rupture appears to be related to operational context.There was no leak noted during preparation or during first inflation, which suggests a product quality issue did not contribute to the reported difficulties.In this case, it is likely that the balloon interacted with the mildly tortuous and mildly calcified anatomy resulting in the reported balloon rupture during the second inflation.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|