ALLERGAN (COSTA RICA) INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SCX-580
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 03/13/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been/will be requested.No additional information is available at this time.The reason(s) for reoperation is/are: capsular contracture baker grade iv.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.
 
Event or Problem Description
Healthcare provider reported a left side capsular contracture baker grade iv.Healthcare provider noted "damage caused by explant" of "yes-side ruptured when removed"(not device related).The device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: the device related to the reported events of rupture and capsular contracture was received on june 05, 2025, with lot number 3242983.Based on the product analysis performed, the assessments of the complaints are: rupture: observed broken device assessed as surgical damage and missing piece of shell assessed as inconclusive.Capsular contracture: unable to observe as it is not related to the manufacturing process.As per the investigation procedures, creases and wear abrasion were completed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
 
Event or Problem Description
Healthcare provider reported a left side capsular contracture baker grade iv.Healthcare provider noted "damage caused by explant" of "yes-side ruptured when removed"(not device related).The device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key21930447
Report Number9617229-2025-07032
Device Sequence Number18410367
Product Code FTR
UDI-Device Identifier10888628030374
UDI-Public(01)10888628030374(10)3242983(11)180824(17)230724
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/24/2023
Device Catalogue NumberSCX-580
Device Lot Number3242983
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/03/2025
Supplement Date Received by Manufacturer06/28/2025
Initial Report FDA Received Date04/29/2025
Supplement Report FDA Received Date07/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/24/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
Patient SexFemale
Patient Weight51 KG
-
-