MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SMHX255
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hematoma (1884)
Date of Event 03/03/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent a breast augmentation revision surgery with implantation of a 255cc mentor memorygel xtra breast implant prosthesis.Post-operatively, the patient was diagnosed with a right breast hematoma using ultrasound imaging.As a result, the patient underwent right-sided removal and replacement with a 255cc mentor memorygel xtra breast implant on (b)(6) 2025.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: hematoma.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On may 08, 2025, mentor completed an evaluation on an image that was provided of the suspect medical device.Upon visual evaluation of the image provided in the complaint no apparent damage or visual anomalies were observed.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On july 02, 2025, the mentor analysis lab received the suspect medical device for evaluation.Paperwork received with the device provided a patient identifier.Patient identifier: (b)(6).On july 06, 2025, mentor completed an evaluation on the returned device.Device evaluation: mentor conducted a visual inspection of the device.During visual evaluation, no apparent damage or visual anomalies were observed on the breast implant device.Hematoma is a collection of blood within the space around the implant.Symptoms from hematoma may include swelling, pain, and bruising.If a hematoma occurs, it will most likely occur soon after surgery; however, it can occur at any time, such as after an injury to the breast.Small hematomas may be absorbed by the body, while others may require surgery for drainage.Hematoma is a known complication associated with this procedure and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL XTRA BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key21931411
Report Number1645337-2025-04599
Device Sequence Number20995453
Product Code FTR
UDI-Device Identifier10081317027802
UDI-Public(01)10081317027802(17)SMHX255(10)9983110
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberSMHX255
Device Lot Number9983110
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/09/2025
Supplement Date Received by Manufacturer05/08/2025
07/02/2025
Initial Report FDA Received Date04/29/2025
Supplement Report FDA Received Date05/22/2025
07/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-