| Model Number |
MMT-431AH |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial and final mdr (b)(4)- device 2 of 2 e1:patient city: (b)(6) patient country: canada.
|
| |
|
Event or Problem Description
|
|
Reference number (b)(4).Event occurred in canada it was reported that the patient faced insulin flow blocked alarm event on 1-apr-2025.The blockage was in the tubing.The patient replaced the infusion set and resumed insulin deliveries successfully.No further information available.
|
| |
|
Additional Manufacturer Narrative
|
|
Supplemental report 01 - (b)(4) - mdr 8021545-2025-00873.The initial mdr with manufacturing report number (8021545-2025-00873), was submitted on 29-apr-2025.Upon completion of the investigation, the manufacturing date was updated as 13-mar-2024.Additional information - this mdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary - complaint investigation results: a complaint investigation has been initiated under complaint investigation child record (b)(4).The batch 6005783.In question was manufactured at the osted site.Device history record (dhr) review: the 6005783 was manufactured according to the work instruction (wi) [79] appendix 1 batchcard for production of packaging room on 13-mar-2024 with a total of (b)(4) units.Review of the dhr showed that all relevant tests required during the related processes had been fulfilled and met the requirements.No deviation was identified, nor maintenance events were recorded related to complaint code.Test results: no product was returned.Conclusion summary of complaint investigation: as a result of the following, no non-conformance (nc) raised during production related to complaint code, no trend identified for the lot in question, no further actions are required.This complaint will not require further root cause investigation nor corrective and preventive action (capa) plan.Therefore, this issue will be monitored through the post market surveillance activities.
|
| |
|
Event or Problem Description
|
|
To date no additional patient or event details have been received.
|
| |
|