• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMP,TSV,4.7,16,MTX,MG; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMP,TSV,4.7,16,MTX,MG; DENTAL IMPLANT Back to Search Results
Catalog Number TSVTWB16
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Osteolysis (2377)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).Multiple reports are being submitted for this event.B3: date of event unknown / not provided.G4: pma/510(k) number: k101880.
 
Event or Problem Description
It was reported that the implants at tooth sites #21 & 28 were removed due to bone loss.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report b5: describe event or problem d9: device availability g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h6: adverse event problem h10: additional narrative dhr review was completed for the subject lot number 1227813.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1227813 for similar events and 1 other relevant complaint(s) was identified.Review completed utilizing keywords: dental: medical: bone loss.A summary investigation has been completed for bone loss events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP,TSV,4.7,16,MTX,MG
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key21932984
Report Number0002023141-2025-01135
Device Sequence Number8796276
Product Code DZE
UDI-Device Identifier00889024020023
UDI-Public(01)00889024020023(17)240518(10)1227813
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K101977
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/17/2024
Device Catalogue NumberTSVTWB16
Device Lot Number1227813
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/30/2025
Supplement Report FDA Received Date07/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/19/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexMale
Patient Weight84 KG
Patient EthnicityNon Hispanic
-
-