• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. J-VAC RESERVIOR BULB 100 CC; CATHETER TIP SUCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. J-VAC RESERVIOR BULB 100 CC; CATHETER TIP SUCTION Back to Search Results
Catalog Number 2160
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint#: (b)(4).Additional information has been requested and received.Attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.Was the reservoir used with the adapter included with the corresponding blake drain? unk.Was the issue noticed before activation? yes.Please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.The device has been received at sukagawa and will be shipped.Please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep or loc/bq) (b)(6).H3 evaluation: complaint sample review: one complaint sample of reservoir bulb with separated connection port was received for evaluation, product was inspected visually: 1.Connection port of the bulb was separated, and there was a cut mark visible on the remaining portion of the bulb port.2.Separated part of the connection port also having visible cut marks.Based on observations, it is suspected that the product might have handled differently at user end, and lead to the defect generation.As per standard practice, 100% functional test and 100% visual inspection was carried out, before and after packing of finished goods, prior to the product release.There was no scope to miss such defect, at manufacturing / release stage.Documents review : during investigation of complaint, batch review of in-process and final packaging inspections, as well as the manufacturing process is performed, these products are reported to be manufactured, inspected, and packaged in conformance with customer's specifications and have been found to be acceptable within all parameters.No discrepancy as per the reported defect is observed.Retention sample review : no negative observation was found.Review of defective sample's image / video : not applicable, as there was no complaint sample image / video received for evaluation.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported a patient underwent an unknown surgery on an unknown date and a bulb reservoir was used.It was broken when inserting to the connection part.There was no effect on the patient.Another product was used to complete the case.There were no adverse consequences to the patient.Further details are not provided.Additional information has been requested.Upon receipt and analysis, it was noted the connection port of the bulb was separated, and there was a cut mark visible on the remaining portion of the bulb port.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
J-VAC RESERVIOR BULB 100 CC
Common Device Name
CATHETER TIP SUCTION
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
DEGANIA INDIA
251, sector-6, imt manesar
gurugram 12205 2
IN   122052
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key21933769
Report Number2210968-2025-04708
Device Sequence Number18369507
Product Code JOL
UDI-Device Identifier10705031003293
UDI-Public(01)10705031003293(11)240214(17)290214(10)J2400068
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
CL II EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2160
Device Lot NumberJ2400068
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2025
Initial Report FDA Received Date04/30/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-