DEPUY SYNTHES PRODUCTS LLC BEARING SLEEVE, 15CM CURVED; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
|
Back to Search Results |
|
| Catalog Number |
MA-15C |
| Medical Device Problem Codes |
Device Reprocessing Problem (1091); Noise, Audible (3273); Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was noted in the service order that the bearing sleeve device had black dirt emerging from it and was producing an abnormal noise.During service and evaluation, it was determined that the device generated heat.It was further determined that the device had a worn bearing, component damage and vibration.It was observed that the device produced an unexpected noise.It was further determined that the device failed pretest for visual assessment and temperature assessment.This event did not occur during surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2025.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11: additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device generating heat, identified during service and evaluation was confirmed.The assignable root cause was determined to be due to component failure from wear.
|
| |
|
Search Alerts/Recalls
|
|
|