| Catalog Number |
254511182 |
| Medical Device Problem Code |
Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : according to the information received "it was reported that the screw backed out on drill.Sleeve(stop) will not slide over drill.¿ the product was not returned to medtech orthopedics ; however, photos were provided for review.(b)(4).The device associated with this report was not returned to medtech orthopedics for evaluation.The photographs attached were reviewed, however in the images provided no observations pertaining to the nature of the reported event of will not seat or advance could be identified.Functionality issues cannot be evaluated through a photo investigation.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was unconfirmed as the photographs attached don¿t have any evidence to draw a conclusion about the reported event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the screw backed out on drill at the collar.Sleeve (stop) will not slide over drill.
|
| |
|
Search Alerts/Recalls
|