MENTOR TEXAS MENTOR MEMORYGEL BOOST BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SMPB590
Medical Device Problem Code Material Discolored (1170)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 01/21/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient implanted with a 590cc mentor memorygel boost breast implant experienced capsular contracture on the right side post-operatively.As a result, the patient underwent explantation on (b)(6) 2025.Upon explantation, the device was found to be discolored.
 
Additional Manufacturer Narrative
Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted visual inspection and leak testing of the returned device.During visual evaluation of the smo ro mod pro boost gel 590cc returned device, the gel was observed to be white.Leak testing was performed, in accordance with mentor procedures, and no gel exposure were detected during the analysis.Discoloration of the implants as observed in the samples returned is related to the concentration of the triglycerides in the gel fillers and their degree of unsaturation.This finding is consistent with the reported discoloration of breast implant silicone gel in vivo in the scientific literature.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet the american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsule, and effusion to pathology for examination.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BOOST BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key21935431
Report Number1645337-2025-04621
Device Sequence Number20804228
Product Code FTR
UDI-Device Identifier00081317026945
UDI-Public(01)00081317026945(17)270224(10)9698840
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSMPB590
Device Lot Number9698840
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/09/2025
Supplement Date Received by Manufacturer05/23/2025
Initial Report FDA Received Date04/30/2025
Supplement Report FDA Received Date05/27/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-