• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC VERSE CORRECTION KEY; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC VERSE CORRECTION KEY; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 199721000
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/10/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: product investigation summary: the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found that the external threads of the verse correction key was heavily striped due to the implantation attempts.Additionally the head of the internal screw was deformed.The observed condition of the device is consistent with a component failure that was caused by exposure to unintended forces like overtightening the device while assemblance; or by assembling the device in a misaligned position.Properly handling and attention to the approved use of the device diminishes the risk of failure.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test was unable to be performed due to the post-manufacturing damage.The overall complaint was confirmed as the observed condition of the verse correction key was consistent with the reported condition.Based on the investigation findings, the potential cause is traced to component failure, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h4, h6: a manufacturing record evaluation was performed for the finished device product code: 199721000 lot number: 353332 it was electronically reviewed and no nonconformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: 05 sep 2022 manufacturing site: jabil le locle if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the surgeon had difficulty tightening the expedium verse correction key during the spine procedure.When the surgeon tried to tighten it, the instrument slipped, making it impossible to tighten the correction key.However, when the correction key was replaced with a new one, the surgeon had no issues and was able to tighten it successfully.The procedure was completed successfully with no surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSE CORRECTION KEY
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key21936076
Report Number1526439-2025-00557
Device Sequence Number17481768
Product Code NKB
UDI-Device Identifier10705034466132
UDI-Public(01)10705034466132(10)353332
Combination Product (Y/N)N
Initial Reporter CountryBX
PMA/510(K) Number
K200245
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number199721000
Device Lot Number353332
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/08/2025
Initial Report FDA Received Date04/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-