| Model Number |
VICM5 13.2 (-8.00) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
04/16/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6- health effect- clinical code: 4581- significant reduction of iridocorneal angles.H11- manufacturer narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced an excessive vault and significant reduction of iridocorneal angles.In a separate visit the lens was exchanged for a shorter length lens and the problem was resolved.The cause of the event was reported as other.Cause details provided: "size problem we need new size between 13.2 and 12.6".
|
| |
|