MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INNOVA? VASCULAR; STENT, SUPERFICIAL FEMORAL ARTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INNOVA? VASCULAR; STENT, SUPERFICIAL FEMORAL ARTERY Back to Search Results
Model Number H74939293061530
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that partial deployment occurred.The limited stenosed target lesion was located in the mildly tortuous and mildly calcified peripheral artery.A 6 x 150 x 130 innova vascular self expanding stent was selected for use.During deployment, the third click sounded different and the stent failed to deploy any further.The stent and catheter shaft were slowly and carefully removed from the body.The stent catheter was placed under fluoroscopy to ensure that the stent was still on the deployment shaft.The stent was safely removed.The procedure was completed with another of the same device.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA? VASCULAR
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key21936276
Report Number2124215-2025-24392
Device Sequence Number12735807
Product Code NIP
UDI-Device Identifier08714729874041
UDI-Public(01)08714729874041(17)290514(10)34035112
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P140028
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH74939293061530
Device Catalogue NumberH74939293061530
Device Lot Number0034035112
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2025
Initial Report FDA Received Date04/30/2025
Date Device Manufactured05/15/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-