| Model Number |
TFNT40 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Visual Impairment (2138)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A healthcare professional reported that, after intraocular lens implantation surgery, the patient experienced continued discomfort.Hence the lens was explanted and replaced with another lens in a secondary procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in d.9., h.3., h.6.And h.11.The lens was returned inside a cylindrical tube.A blue lid (r) was on one end of this container and a pink lid (l) was on the opposite end.The lens was removed to evaluate.The optic was cut into two pieces, typical of an insertion and removal.One optic edge was broken.The broken area was not returned.The posterior optic surface was cracked near the edge.Product history records were reviewed and the documentation indicated the product met release criteria.Associated product information was not provided.The product investigation could not identify a root cause for the reported complaint.The lens was returned cut into pieces, typical of an insertion and removal.The optic had additional damage.Each lens is subjected to a 100 percent assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.The power and resolution of the returned lens could not be re-evaluated due to the optic damage.Information was provided that the (2.25cyl), 06.0 diopter (add power +2.2/+3.2) lens was replaced with an unspecified lens model.Additional information was provided the initial lens was implanted on (b)(6) 2020 and explanted on (b)(6) 2020.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|