(b)(4).H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.The following information was requested, but unavailable: were there any adverse consequences associated with the patient? did the needle fall into the patient? if so, was the needle piece(s) retrieved during the same procedure? were x-rays taken to locate the needle piece(s)? what measures were implemented to retrieve the broken piece? was there any additional tissue damage resulting from the search for the needle piece? received information as following from sales rep via phone today.Please refer to the event description, other information requested is unknown.H3 investigation summary: this is an analysis for a photo submitted to ethicon for evaluation.During the visual analysis, the following was observed: the photo shows an apparent detachment of the needle.A clear analysis of the end point of failure or the needle hole could not be performed due to the position of the needle.Based on the photo review, the event describe is confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the device was not returned our evaluation is limited.We value the opportunity to fully analyze the device upon its return.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.
|
|
It was reported that a patient underwent a laparoscopic ureterolithotomy surgery on (b)(6) 2025 and suture was used.During the procedure, the patient underwent surgery on the morning.During the laparoscopic ureterolithotomy surgery, when the surgeon was suturing the skin, the tail of the needle suddenly broke, and the surgery was immediately stopped.After two people checked, it was confirmed that the needle body was complete.Immediately replaced the suture with a new one for suturing and used normally.No adverse patient consequences were reported.Additional information was requested.
|