Section a2, a3, a4 and a5: per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section d6b: explant date: not applicable, as lens remains implanted.Section e1: telephone number: (b)(6).Device evaluation: product testing could not be performed since the device was not returned for evaluation.Therefore, the reported issue could not be verified, and no product quality deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order (po) was performed.The search revealed that no additional complaints were received for this po.No escalation required.Conclusion: based on the complaint investigation results, no product deficiency or product malfunction were found.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
|
It was reported that during the implantation of the preloaded monofocal intraocular lens (iol), a small piece of plastic followed the lens into the patient's eye.The iol itself was not damaged, and the plastic was successfully removed.The patient's eye was not affected by this incident.The patient has fully recovered.Through follow up, we learned that there is no material available for return.No further information was provided.
|