|
MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3501751BC |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
02/24/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture.D4: udi: as the serial number for the device involved in the event was not provided, the full udi is currently not available.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient underwent a breast augmentation with two 175cc mentor memorygel breast implants and experienced bilateral capsular contracture (baker grade 3) post-operatively.As a result, the patient underwent bilateral explantation on (b)(6) 2025, and replaced with two 225cc mentor memorygel breast implant (catalog number 3502254bc) with serial number (b)(6), on the left and (b)(6), on the right.This report is for the left side device.
|
| |
|
Additional Manufacturer Narrative
|
|
On may 08, 2025, a photo evaluation for the device was completed.Photo evaluation summary: during visual evaluation of the image provided, some bubbles were observed in the gel.As the product involved in this complaint was not received, the device could not be analyzed according to our procedures.Mentor also performed a manufacturing record evaluation related to the reported complaint for the finished device lot number.No manufacturing non-conformances were identified as part of this evaluation.The shell wall of the implant is permeable to air, and trapped air can form bubbles with changes in pressure or temperature.When left by themselves with no changes in temperature or pressure, bubbles usually dissipate over time unless trapped by cured gel.Based on customer feedback and our complaints database, discrete bubbles do not persist in implanted devices, nor are they found in previously implanted devices.The presence of these bubbles is not known to compromise the performance of the device.Through our post-market surveillance process, we have not identified any patient harm associated with bubbles within implanted devices.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet as part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On july 02, 2025, the mentor failure analysis lab has received the device for evaluation.The patient identifier was updated to (b)(6).The patient's date of birth was reported to be (b)(6) 1983.The patient's initial procedure was reported to be a breast augmentation revision.On july 05, 2025, the evaluation for the device was completed.Device evaluation summary: visual analysis of the returned device revealed that the breast implant was found to have some bubbles within the gel.Leak testing was performed, in accordance with mentor procedures, and no areas of gel exposure were detected during the analysis.The shell wall of the implant is permeable to air, and trapped air can form bubbles with changes in pressure or temperature.When left by themselves with no changes in temperature or pressure, bubbles usually dissipate over time unless trapped by cured gel.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|