It was reported that an inserter broke during a spinal surgical procedure.After the implant was placed in the disc space, the inserter was removed.Upon removal, it was observed that the portion of the tip that locks into the implant had fractured.The broken tip was retrieved without complication.
|
|
Correction: h3.Yes h6.Investigation findings: 3251 investigation conclusions: 22.Device evaluation: visual inspection confirms that the arm component has fractured.The arm of the inserter is a subcomponent which features an ipsilateral prong that attaches to the interbody spacer for insertion.This type of damage occurs when force is applied during impaction/insertion of the interbody spacer causing the distal arm to detach from the shaft.Review of device history records showed no manufacturing or processing related irregularities that would cause or contribute to this failure mode.Labeling review: warnings/cautions/precautions: potential risks identified with the use of these fusion devices, which may require additional surgery, include device component failure, loss of fixation, pseudoarthrosis (i.E.Non-union), fracture of the vertebra, neurological injury, and/or vascular or visceral injury.Possible adverse effects: initial or delayed loosening, bending, dislocation, and/or breakage of device components.
|