• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC ENDOWRIST; ROUND TIP SCISSORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC ENDOWRIST; ROUND TIP SCISSORS Back to Search Results
Model Number 470007-05
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that after a da vinci-assisted surgical procedure, the 8mm round tip scissors instrument exhibited a small fracture of the cable.No fragments fell into the patient.The procedure was completed with no reported injury.
 
Additional Manufacturer Narrative
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
 
Event or Problem Description
Refer to h11 for follow-up information.
 
Additional Manufacturer Narrative
The instrument involved in this complaint has been received and tested by failure analysis.The instrument was found to have a damaged pitch cable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Common Device Name
ROUND TIP SCISSORS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
4085232100
MDR Report Key21938793
Report Number2955842-2025-17757
Device Sequence Number5355607
Product Code NAY
UDI-Device Identifier00886874112168
UDI-Public(01)00886874112168(11)231219(10)K10231219(91)0001
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K150284
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Notification,Other
Type of Report Initial,Followup
Report Date (Section B) 04/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number470007-05
Device Catalogue Number470007
Device Lot NumberK10231219 0058
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/08/2025
Supplement Date Received by Manufacturer05/02/2025
Initial Report FDA Received Date04/30/2025
Supplement Report FDA Received Date05/22/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/19/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
ISIFA2024-09-C
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES.
Patient SexUnknown
-
-