| Model Number |
VTICMO 13.2 (-16.0/2.5) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
10/30/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation is required but has not yet been performed.H6 - investigation type 4110: lens work order search - no similar complaint event(s) within associated lots were found.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim #: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise, excessive vault, and glares/haloes.The lens was explanted and replaced with a shorter length lens with a different diopter.Cause was reported as other.Reportedly, "excessive vault.".
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation - lens was returned dry in liquid in vial.Visual inspection found haptic torn.Dimensional and functional inspection found the lens to be within specifications.Claim#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: e1 - initial reporter: (b)(6).Claim#: (b)(4).
|
| |
|