STRYKER ORTHOPAEDICS-MAHWAH TRITANIUM BPLATE TRIATHLON S6; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5536B600
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/16/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient's right knee was revised due to the posterior medial third of the baseplate and the insert breaking off.
 
Additional Manufacturer Narrative
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event or Problem Description
No new information.
 
Additional Manufacturer Narrative
An event regarding crack/fracture involving a triathlon baseplate was reported.The event was confirmed via inspection of the returned devices.Method & results: -device evaluation and results: visual inspection & material analysis: review of the tritanium tibial baseplate, catalogue # 5536-b-600, lot code ctd51601 and tibial insert, catalogue # 5531-g-012-e, lot code t93152 confirmed fracture of the device.Characterisation using stereo microscopy confirmed the presence of embedded debris on both the articulating and distal surfaces of the insert.Characterisation using stereo microscopy and scanning electron microscopy confirmed fracture through the baseplate in fatigue.Characterisation using stereo microscopy confirmed the fracture of the insert and evidence of embedded debris on both the articulating and distal surfaces of the insert.No manufacturing or material related defects were observed on the examined areas of the components.-clinician review: no medical records were received for review with a clinical consultant.-device history review: review of the product history records indicate the reported device was manufactured with no reported relevant discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to fractured devices.Visual inspection & material analysis: review of the tritanium tibial baseplate, catalogue # 5536-b-600, lot code ctd51601 and tibial insert, catalogue # 5531-g-012-e, lot code t93152 confirmed fracture of the device.Characterisation using stereo microscopy confirmed the presence of embedded debris on both the articulating and distal surfaces of the insert.Characterisation using stereo microscopy and scanning electron microscopy confirmed fracture through the baseplate in fatigue.Characterisation using stereo microscopy confirmed the fracture of the insert and evidence of embedded debris on both the articulating and distal surfaces of the insert.No manufacturing or material related defects were observed on the examined areas of the components.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRITANIUM BPLATE TRIATHLON S6
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NJ NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key21940909
Report Number0002249697-2025-00456
Device Sequence Number9055880
Product Code MBH
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/03/2025
Device Catalogue Number5536B600
Device Lot NumberCTD51601
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/16/2025
Supplement Date Received by Manufacturer05/11/2026
Initial Report FDA Received Date05/01/2025
Supplement Report FDA Received Date06/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/22/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
-
-