• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TLOC 133 MP SP T1 PPS HO 7X99; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TLOC 133 MP SP T1 PPS HO 7X99; PROSTHESIS, HIP Back to Search Results
Catalog Number 51-109070
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the while inspecting the product, the packaging was found damaged.There was no patient involvement.There is no further information available.
 
Additional Manufacturer Narrative
(b)(4).D10: 51-102040 item name tprlc xr fp type1 pps 4x128mm lot # 6022932 51-109070 item name tloc 133 mp sp t1 pps ho 7x99 lot # 6576225 51-107140 item name tprlc 133 mp type1 pps ho 14.0 lot # 7208588 51-103160 item name tprlc 133 t1 pps so 16x152mm lot # 6095766 g2: foreign: japan the product has been received by zimmer biomet.The investigation is currently in process.Once the investigation has been completed, a follow-up report will be submitted.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (pouch).Sterility had been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The reported event was not related to a combination of products; therefore, a compatibility review is not applicable.The reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.The reported event has been confirmed by evaluation of the returned product and provided photos.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TLOC 133 MP SP T1 PPS HO 7X99
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21941321
Report Number0001825034-2025-01275
Device Sequence Number8800205
Product Code KWA
UDI-Device Identifier00880304513013
UDI-Public(01)00880304513013(17)291209(10)6695289
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K110400
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number51-109070
Device Lot Number6695289
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2025
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2025
Supplement Date Received by Manufacturer08/13/2025
Initial Report FDA Received Date05/01/2025
Supplement Report FDA Received Date08/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/09/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PLEASE SEE H11
Patient SexUnknown
-
-