• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. CONTOUR CURVED CUTTER STAPLER; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. CONTOUR CURVED CUTTER STAPLER; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number CS40G
Medical Device Problem Codes Failure to Form Staple (2579); Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent 5/1/2025.D4 batch # unk.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.No lot or batch number was provided therefore a device history could not be done.Additional information was requested, and the following was obtained: 1.Did the staples protruded before firing? it¿ was not sure.2.Did the device cut? some of part was cut.3.If the device cut/fired, was the cut line complete? yes.4.Did the device staple? yes.5.If the device stapled/fired, was the staple line complete? no.6.If the device was stapled, was the shape of the staples (b-formation, irregular, unformed)? irregular formation.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unknown procedure when surgeon tried to use it, some of stapler legs projected from cartridge deck.There was a surgical delay.The procedure was successfully completed.There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Date sent 7/10/2025.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
 
Additional Manufacturer Narrative
(b)(4).Date sent 8/5/2025.D4 batch #809c18.Corrected data d4: udi# investigation summary- the product was returned for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual analysis of the returned sample determined that the cs40g device was received with no apparent damage and with a reload loaded in the device.The reload was received with some staples protruding, the washer uncut and with the knife recessed below the reload deck.The drivers were noted to be partially advance.The device was tested for functionality with a test reload and the device fired, forming all staples and cutting as intended.The cut line and the staple line were completed and the staples meet the staple form release criteria.The condition of the reload is consistent with the device being partially activated.As a result of the partial firing the lockout was engaged when the device was reopened.Do not fire the instrument unless the closure trigger is properly latched against the handle.The firing trigger must be pulled back completely against the closure trigger to properly fire the instrument.In addition, if the firing sequence is not complete, you could deploy the staples without cutting the washer and forming the staples.Please reference instructions for use for additional information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 809c18, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR CURVED CUTTER STAPLER
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
3035526892
MDR Report Key21941549
Report Number3005075853-2025-03166
Device Sequence Number21032566
Product Code GDW
UDI-Device Identifier10705036012405
UDI-Public(01)10705036012405(17)291231(10)809C18
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K040038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCS40G
Device Lot Number809C18
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/27/2025
Supplement Date Received by Manufacturer07/09/2025
07/25/2025
Initial Report FDA Received Date05/01/2025
Supplement Report FDA Received Date07/10/2025
08/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/25/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-