| Catalog Number |
9770131R |
| Medical Device Problem Code |
Data Problem (3196)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the sherlock device is not working correctly (no p wave elevation).Customer has needed to cxr.No other information provided, at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The manufacturer has received the sample and is pending evaluation.A supplemental will be submitted with evaluation results.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device was returned to service facility for evaluation.During evaluation, the reported issue of the sensor is giving inaccurate readings was unconfirmed.The sherlock sensor was connected to a test sr8 and both the magnet tracking and ecg function properly.There is no root cause as the issue could not be reproduced.
|
| |
|