| Catalog Number |
9770352 |
| Medical Device Problem Codes |
Break (1069); Data Problem (3196)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The manufacturer has received the sample and is pending evaluation.A supplemental will be submitted with evaluation results.
|
| |
|
Event or Problem Description
|
|
It was reported by a tm that the unit is intermittently giving incorrect picc placement.No other information provided, at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device was returned to the service facility for evaluation.During evaluation, the reported issue of unit intermittently giving incorrect placement for picc placement was confirmed.The sherlock sensor does not function when connected to an sr8 test fixture used as a display.The sherlock does not show the lollipop during magnet tracking functionality testing.The reported issue root cause is material failure of the usb cable due to device use.
|
| |
|
Search Alerts/Recalls
|