Updated fields: d4, d9, g3, g6, h2, h3, h4, h11.The duraseal dural sealant system 5ml us box of 5 (202050) was not returned for analysis and the investigation could not confirm the complaint.A dhr review and trending were performed as part of the evaluation.Proper finished good testing was performed prior to release as indicated in the dhr.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.Per the fmea, potential causes of failure include: issues with performance.The risk remains acceptable per the risk analysis.No enhancements or improvements were generated for the reported condition.The file will be closed as could not be reliably determined.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|