| Catalog Number |
XCR120610 |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer ref#: (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section d2b: procode: krd/hcg.Section h3: the device is available to be returned for evaluation and testing.However, it has not been received to date.If the device returns, a device investigation will be performed.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
As reported by a healthcare professional, during a coil embolization of the left middle cerebral artery (mca), a freeform xsft 6mm x 10 cm coil (xcr120610, 31173119) failed to detach.A mechanical detachment handle (mdh1, 102109430) failed.The coil was removed without incident.Replaced with a cerepak ff xsft 4x10 coil.There was no patient injury reported.Additional event information received on 29-apr-2025 indicated that a sl-10 microcatheter (mc) was used.Two coils were previously implanted.The microcatheter was re positioned.The coil was being placed in a tight area with some microcatheter kick back.There was no problem with advancement in the microcatheter.Two attempts were made using the detachment handle.There was no visible damage to the embolic coil.The vessel was not excessively tortuous or with acute bends.The physician mentioned that the force used to get the coil into place along with the microcatheter kick back could have been a factor in the failure to detach.The procedure was delayed by 3 to 4 minutes, with no clinical significance.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer ref#:(b)(4).Updated sections on this medwatch: b4, d9, g3, g6, h2, h3 and h11.The product has been returned for evaluation and testing; however, the engineering evaluation has not been completed.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer ref#: (b)(4) updated sections on this medwatch: b4, g3, g6, h2, h3, h6 and h11.Complaint conclusion: as reported by a healthcare professional, during a coil embolization of the left middle cerebral artery (mca), a freeform xsft 6mm x 10 cm coil (xcr120610, 31173119) failed to detach.A mechanical detachment handle (mdh1, 102109430) failed.The coil was removed without incident.Replaced with a cerepak ff xsft 4x10 coil.There was no patient injury reported.Additional event information received on 29-apr-2025 indicated that a sl-10 microcatheter (mc) was used.Two coils were previously implanted.The microcatheter was re positioned.The coil was being placed in a tight area with some microcatheter kick back.There was no problem with advancement in the microcatheter.Two attempts were made using the detachment handle.There was no visible damage to the embolic coil.The vessel was not excessively tortuous or with acute bends.The physician mentioned that the force used to get the coil into place along with the microcatheter kick back could have been a factor in the failure to detach.The procedure was delayed by 3 to 4 minutes, with no clinical significance.A non-sterile freeform xsft 6mm x 10 cm coil was received contained in the decontamination pouch.Upon receiving the device, visual inspection was performed, and the manual break was attempted at the proper location; however, the embolic coil was still attached to the delivery unit.The pull wire was not exposed at the proximal end.The delivery tube was found kinked at 2.3cm from the proximal key engagement.The embolic coil was noted in an unraveled condition.Microscopic inspection on the embolic coil revealed a kink.No damages were noted on the embolic coil and proximal key engagement, nor unintentional weld was noted at the loop-hole.A manufacturing record evaluation was performed for the finished device 31173119 number, and no non-conformances related to the malfunction were identified.The issue reported regarding the failure to detach is confirmed based on the attached condition of the embolic coil to the delivery unit after activating the manual break.It is suggested that the kinked and unraveled conditions of the embolic coil could be the result of the microcatheter kick back, rendering the device unusable since according to risk documentation, accidental mechanical product damage is a potential failure mode that may result in potential degradation of device performance; failure to detach, and needing to replace the device.There is no indication that the issues reported in the complaint result from a defect inherently related to the cerepak device.As part of j&j medtech quality process all devices are manufactured, inspected, and released to approved specifications.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no capa activity is required.The instructions for use (ifu) contain the following recommendations: 1.Remove the detacher from the protective packaging and place it within the sterile field.The detacher is packaged separately for single patient use only.2.Confirm again under fluoroscopy that the coil alignment marker of the delivery tube creates a ¿t¿ with the proximal marker of the microcatheter.3.Verify that the rhv is firmly locked around the delivery tube to ensure that the coil does not move during connection process.Ensure the delivery tube is straight between the rhv and the detacher.4.Hold the delivery tube approximately 3 inches from proximal end and bring the detacher over the delivery tube until a firm stop is felt.5.Maintain the delivery tube position within the detacher and use a thumb to pull back the slider on the detacher.6.Continue to pull on the slider until end of travel or a click is heard.Gently release the slider until it returns to starting position or a second click is heard.Remove detacher by slowly withdrawing it from the delivery tube.Note: detacher must remain in sterile field until the procedure is completed.7.Fluoroscopically verify that the coil is detached by gently pulling back the delivery tube approximately 1cm.If the coil has detached, remove the delivery tube from the microcatheter.8.If the coil has not detached, repeat steps 3-7.If 2 unsuccessful detachments occur, discard the detacher and proceed to the manual break section.9.Repeat steps 1-8 until the procedure is completed.Discard the detacher.Caution: always perform fluoroscopic verification of detachment prior to withdrawing the delivery system fully.Failure to do so may lead to an embolic complication.10.After detaching the coil, remove the delivery tube from the microcatheter and discard.11.Repeat the above sequence for all additional coils until the procedure is complete.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|