| Catalog Number |
21-2770-24 |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "it was impossible to purge the needle." the nurses changed the needle with one of the same lot number.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for evaluation in used condition.Visual and functional testing was performed, and priming was unsuccessful.The occlusion area was severed at the tubing-needle junction point.A solvent membrane was observed within the needle lumen.The reported issue was confirmed.The probable cause is due to errant solvent applied at the tubing-needle bond.The lot history was reviewed, no nonconformities were identified that may have contributed to the reported complaint.
|
| |
|
Search Alerts/Recalls
|