• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; implanted subcortical electrical Stimulator (motor disorders)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; implanted subcortical electrical Stimulator (motor disorders) Back to Search Results
Model Number 37601
Medical Device Problem Codes Human-Device Interface Problem (2949); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient decided they did not want deep brain stimulation and requested the removal of the implantable pulse generator (ipg) and leads.The leads were capped with boots instead of the correct end caps due to short notice and unavailability.This impacted future magnetic resonance imaging (mri) eligibility unless the entire system is removed or correct end caps are applied.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: product id: 3387s-40, serial# (b)(6), implanted: (b)(6) 2009, product type lead.Section d information references the main component of the system.Other relevant device(s) are: product id: 3387s-40, serial/lot #: (b)(6).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Common Device Name
implanted subcortical electrical Stimulator (motor disorders)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key21943324
Report Number3004209178-2025-07556
Device Sequence Number18342978
Product Code MRU
UDI-Device Identifier00643169529786
UDI-Public00643169529786
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
H020007
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/28/2017
Device Model Number37601
Device Catalogue Number37601
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2025
Initial Report FDA Received Date05/01/2025
Date Device Manufactured04/07/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H11...."
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-