|
H3: product analysis #821542297:¿ equipment used: vis (m-78210) ¿ drawing(s) referenced: none ¿ as found condition: the mirage guidewire was returned for analysis within a shipping box and a plastic bio-pouch.¿ damage location details: the distal end of the mirage guidewire was found bent, with the outer coils stretched and the inner core wire broken.The mirage guidewire distal tip was found bent.The characteristic of the bent distal tip is consistent with shaping.¿ testing/analysis: the broken mirage guidewire was sent out for sem failure analysis.Per the sem report, the broken end failed via rotational overload.¿ conclusion: based on the device analysis and reported information, the customer¿s ¿guidewire damage¿ report was confirmed.The guidewire can become damaged due to patient vessel tortuosity, aggressive removal of the guidewire, use of an incompatible catheter, or advancing or retrieving the device against resistance.From the damage observed on the returned mirage guidewire, force was applied.However, the cause of the guidewire damage could not be determined.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|