It was reported that a patient underwent a bypass surgery on (b)(6) 2025 and a barbed suture was used.During the procedure, the suture had been found broken in the newly dispensed suture when it was opened to prepare for bypass surgery.Changed another one to complete the surgery.There were no adverse patient consequences reported.Upon evaluation of the returned device, short insertion was observed in the strand.Additional information was requested.
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.The following information was requested but unavailable: if possible, could you kindly provide the lot number, please? --unknown.H3 investigation summary: the product was returned to ethicon for evaluation.A detached needle and one suture outside the winding former were received for analysis.Product code sxpp1a410.After investigation of the sample, short insertion was observed in the strand.The visual inspection concluded that the combination of the needle/suture was detached from the needle since the insertion end of the suture did not meet the requirement.Based on the information currently available, short insertion issue was identified as pull out during the investigation of the sample received.This product issue will be addressed through the ethicon quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Device history review- the manufacturing records could not be reviewed as the batch/lot number is unknown.
|