It was reported that the short attachment device bearing was damaged.During in-house engineering evaluation, it was determined that the device was heating up, vibrating, made unexpected noise and was difficult to assemble/disassemble.The device also failed pretests for lock operation, vibration and temperature assessment.There were no adverse patient consequences nor surgical delay reported.The date of event was unknown but was known to have occurred in 2025.No additional information was provided.
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the actual device was returned for evaluation.The device was evaluated and it was determined that the reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, during evaluation, it was determined that the device was heating up, vibrating, made unexpected noise and was difficult to assemble/disassemble.The device also failed pretests for lock operation, vibration and temperature assessments.The assignable root cause was traced to component failure due to wear.
|