| Model Number |
470001-09 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the 8mm potts scissors instrument to perform failure analysis.There was no physical damage.There was no problem detected.The instrument was fully functional.No product issue was found.The complaint regarding broken cable was not confirmed by failure analysis.Based on additional information, it has been determined that the instrument involved with this complaint does not meet the criteria for isifa2024-09-c as previously listed in section h9.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|