| Model Number |
4965-15 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: w1dr01 ipg, implanted on (b)(6) 2024.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead exhibited high impedances, no capture and a polarity switch.A lead impedance warning and polarity switch was also observed on the right atrial (ra) lead.It was noted that the ra lead failed an atrial lead position check.The rv lead was capped and replaced.The ra lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|