Model Number 55750025540
Medical Device Problem Codes Backflow (1064); Leak/Splash (1354)
Health Effect - Clinical Code Extravasation (1842)
Date of Event 04/04/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having l2 posterior sp inal fusion for an indication of th12 vertebral fracture.It was reported that during cement injection into the l2 fsn, the cement was leaked to the blood vessel-like site outside the vertebral body.The cement was confirmed in the in the right atrium in the postoperative ct.There was a delay of less than 60 mins reported. the additional surgery was performed on (b)(6) 2025 and a dry sheath and a snare catheter were inserted through the neck to remove cement that had entered the right atrium.As for the cement around the spine, it is unlikely that it will be removed.There was no patient symptom reported.There were no further complications reported regarding the event.
 
Additional Manufacturer Narrative
A: select patient information cannot be included in regulatory report due to regional privacy regulations.G2: the country of the event is japan.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
B1.Advent/prod prob corrected.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON FENESTRATED SCREW SET
Common Device Name
BONE CEMENT, POSTERIOR SCREW AUGMENTATION
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key21948901
Report Number1030489-2025-01863
Device Sequence Number15633390
Product Code PML
UDI-Device Identifier00763000149574
UDI-Public00763000149574
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K201362
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number55750025540
Device Catalogue Number55750025540
Device Lot NumberH5744966
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2025
Supplement Date Received by Manufacturer05/02/2025
Initial Report FDA Received Date05/02/2025
Supplement Report FDA Received Date05/27/2025
Date Device Manufactured07/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-