ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL; INTRAOCULAR LENS Back to Search Results
Model Number SY60WF
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/03/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non healthcare professional reported that during an intraocular lens (iol) implant procedure, the surgeon noticed haptic was broken after implantation.Lens was removed and new lens was implanted.Additional information has been received that there was no issues or harm to the patient.
 
Additional Manufacturer Narrative
The lens was returned positioned incorrectly in the lens case.Solution was dried on the lens.One haptic was broken from the haptic/optic junction area (not returned).The optic was cut in half, typical of insertion and removal.Product history records were reviewed and the documentation indicated the product met release criteria.Qualified associated products were indicated.The reported broken haptic was observed.All lenses are 100 percent inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The ifu (instructions for use) instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21949126
Report Number1119421-2025-01266
Device Sequence Number8720423
Product Code HQL
UDI-Device Identifier00380652251259
UDI-Public(01)00380652251259(17)280908(11)240910(10)15913101
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSY60WF
Device Catalogue NumberSY60WF.085
Device Lot Number15913101
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/03/2025
Supplement Date Received by Manufacturer05/21/2025
Initial Report FDA Received Date05/02/2025
Supplement Report FDA Received Date06/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/10/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH II LOADING FORCEPS; MONARCH III IOL CARTRIDGE D; MONARCH III IOL INJECTOR; PROVISC OVD
Patient SexUnknown
-
-