| Catalog Number |
EVD35-06-100-120 |
| Medical Device Problem Codes |
Premature Activation (1484); Malposition of Device (2616); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that an inadvertent deployment of a 6 x100 everflex entrust self-expanding stent in the common femoral artery extending into the superficial femoral artery occurred in a patient.The deployment mechanism appeared to break.The patient required hospitalization as a result of this event.
|
| |
|
Search Alerts/Recalls
|