| Catalog Number |
157001090 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Necrosis (1971); Unspecified Tissue Injury (4559)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Depuy orthopaedics is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy orthopaedics has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy orthopaedics or its employees that the report constitutes an admission that the device, depuy orthopaedics, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h3, h6: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a primary surgery with the summit stem and the cup (non-j&j product).The surgery was completed successfully without any surgical delay.After surgery, retroversion of the pinnacle a (metal-on-metal) cup occurred, and armd was suspected.On (b)(6) 2025, the revision surgery was performed.Necrotic tissue spread in the large area of soft tissue were removed.A black ring-shaped foreign object was found attached to the neck part of the summit.When attempting to remove the head, metal liner, and 2 screws, the cup was found to be loose, so they were able to be easily removed by hand.A pinnacle gription multi hole cup 56 mm was fixed with 5 screws using a competitor's navigation.A 36 mm neutral liner was installed, trial heads with 36 mm+1.5 and +5 were used for fixation, and the +5 one was selected eventually.A delta ts head was hammered in, and the revision was completed.
|
| |
|
Search Alerts/Recalls
|
|