MEDTRONIC DOMINICANA CSF-FLOW CONTROL VALVE; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number 42544
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/30/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information regarding a valve.It was reported that the surgeon stated the valve was faulty, and they suspected it to be occluded/blocked.The valve was explanted and revised where a new valve was inserted.The patient was currently recovering, and their status at the time of the report was alive-no injury.
 
Additional Manufacturer Narrative
H3.Product analysis determined that the valve did not meet the requirements for pressure-flow testing.There was debris observed within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open affecting the flow of fluid through the valve.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSF-FLOW CONTROL VALVE
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC DOMINICANA
zona franca de san isidro
carretera san isidro, km 17
santo domingo 11500
DR  11500
Manufacturer (Section G)
MEDTRONIC DOMINICANA
zona franca de san isidro
carretera san isidro, km 17
santo domingo 11500
DR   11500
Manufacturer Contact
justin ellis
4620 n. beach st.
fort worth, TX 76137
7635265677
MDR Report Key21949317
Report Number9612501-2025-01206
Device Sequence Number8358893
Product Code JXG
UDI-Device Identifier00763000028398
UDI-Public00763000028398
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K831678
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number42544
Device Catalogue Number42544
Device Lot Number0227642351
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2025
Supplement Date Received by Manufacturer05/30/2025
Initial Report FDA Received Date05/02/2025
Supplement Report FDA Received Date06/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;