| Model Number |
PNDL, 5BV WGN 32G 4MM 50CT 60/CS |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report reference number: (b)(4).Pen needles were not returned for evaluation.Note: manufacturer contacted customer in several follow-up calls to ensure the replacement products resolved the initial concern - unable to establish contact with customer at this time.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for the trueplus pen needles.Customer stated that the 32g pen needles did not dispense the insulin.The package had not been open or damaged when received.The customer has been using the product for 1 month.The customer is using compatible product, and the pen needle is properly aligned.At the time of the call the customer felt well and did not report any symptoms.Customer did not claim to be injured while using the pen needles and no medical intervention related to the use of the product was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Sections with additional information as of 25-jun-2025: h6: updated fda¿s type, findings and conclusions codes.H10: pen needles were not returned for evaluation.Complaint was forwarded to supplier quality and internal evaluation was performed by the manufacturer using pen needles from the same lot.No abnormalities observed with retention samples.Most likely underlying root cause: mlc-061: improper use/mishandled by end user.
|
| |
|