| Model Number |
VICMO_13.2 (-16.0) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/05/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, a patient experienced excessive vault and significant reduction of irido-corneal angles.The lens was later exchanged for a shorter length lens and the problem was resolved.Cause of the event is a unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: work order search: no similar complaints within associated lots were found.Manufacturer narrative: secondary surgical intervention to remove and replace the lens are identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|