ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA TIBIA CEMENTED 5 DEG STEMMED LEFT SIZE G; PROSTHESIS, KNEE Back to Search Results
Catalog Number 42532007901
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032)
Date of Event 04/10/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).G2: australia.Suggested code (annex g)- mechanical (g04) - femur.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device was requested but not returned by the facility.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.D10: additional associated products 42502607401 persona femur cemented (cr) standard left size 12 lot# 64910165.
 
Event or Problem Description
It was reported the patient underwent a knee revision surgery due to pain from flexion/extension mismatch.The femoral and tibial implants revised.The articular surface also replaced at time of revision.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.Updated: b4, b5, d5, g3, g6, h1, h2, h3, h6, h10, h11.Suggested code (annex g)- mechanical (g04) - femur.Visual examination of the provided photos identified explanted total knee arthroplasty components, including a femoral component, tibial baseplate, and polyethylene insert.All components exhibit biological debris consistent with in vivo use.The polyethylene insert is marked ¿mc left ve¿ with part and lot information visible.The tibial and femoral components show residual cement.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The reported products were reviewed for compatibility with no issues noted.Medical records were not provided.Complaint is not confirmed.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.D10: additional associated products: 42502607401, persona femur cemented (cr) std left size 12, lot# 64910165.42512101011, persona articular surface (mc) left 11 mm, lot# 64764065.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA TIBIA CEMENTED 5 DEG STEMMED LEFT SIZE G
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21951943
Report Number3007963827-2025-00179
Device Sequence Number12970736
Product Code MBH
UDI-Device Identifier00889024471092
UDI-Public(01)00889024471092(17)310414(10)65001961
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number42532007901
Device Lot Number65001961
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2025
Supplement Date Received by Manufacturer05/30/2025
Initial Report FDA Received Date05/02/2025
Supplement Report FDA Received Date06/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-