Blank fields on this form indicate the information is unknown or unavailable.H3: device evaluation anticipated, but not yet begun.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
As reported, during an angiogram and angioplasty of the left leg, a flexor ansel guiding sheath separated while in the patient.The device had been in use for one-to-two hours at the time of separation.The customer reported a ¿near miss.¿ additional information has been requested.
|
|
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Summary of event: as reported, during an angiogram and angioplasty of the left leg, a flexor ansel guiding sheath separated while in the patient.The device had been in use for one-to-two hours at the time of separation.The customer reported a ¿near miss.¿ investigation evaluation: reviews of the complaint history, device history record (dhr), instructions for use (ifu), and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The complaint device was returned to cook for investigation.The sheath was separated into two sections.The proximal segment of the sheath was damaged, and approximately 9.9-centimeters of the sheath remained attached to the hub.The distal sheath segment measured approximately 33.1-centimeters in length.A document-based investigation evaluation was performed.A review of the device history record found no relevant non-conformances on the final or sub-assembly lots.A review of complaint history found no additional complaints for this lot number.The product ifu states ¿if resistance is encountered during advancement of flexor sheath, assess cause of resistance and consider dilation of any restriction identified or consider alternate treatment strategy.If flexor sheath is advanced through resistance, force to remove the sheath will be higher, increasing the risk of sheath material or hub separation upon withdrawal.¿ a review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Based on the information provided and the results of the investigation, cook has concluded that a component failure, unrelated to manufacturing or design deficiencies, contributed to this event.The customer did not provide additional information to cook.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified, and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|