• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ATTUNE REV LPS INSRT XXSM 12MM; KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ATTUNE REV LPS INSRT XXSM 12MM; KNEE TIBIAL INSERT Back to Search Results
Catalog Number 151760112
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Joint Laxity (4526)
Date of Event 04/10/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A manufacturing record evaluation (nc search) was performed for the finished device product code:151760112, lot: m7537m and no non-conformances / manufacturing irregularities were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: a manufacturing record evaluation (nc search) was performed for the finished device product code:151760112, lot: m7537m and no non-conformances / manufacturing irregularities were identified.Added: d10 (concomitant), corrected: d4 (primary udi number).
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.
 
Event or Problem Description
It was reported that the patient presented with poor patellar tracking and was scheduled for surgery to address.After exposure, surgeon noted the femoral component appeared to be internally rotated.While extracting the femoral component he noticed that the femoral sleeve also moved indicating that it had not yet ingrown.This led the surgeon to believe that the patient could still be infected as he was treated at the previous surgery with a one stage revision for infection.Frozen section was sent to pathology and showed 5 white cells/ high powered field.This result was more indicative of a loose component than infection but surgeon felt compelled to proceed as if infected and also sent cultures.Femoral components were removed and a thorough debridement and irrigation was performed.Femur was then prepared for new components and placement of resorbable antibiotic beads.No delay was encountered and no patient harm noted.Patellar tracking and motion were found to be improved with new components in place.There was loosening reported at non cemented interface at femoral sleeve.Doi: (b)(6), 2025 dor: (b)(6) 2025 affected side: left knee.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE REV LPS INSRT XXSM 12MM
Common Device Name
KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key21952618
Report Number1818910-2025-06943
Device Sequence Number8535085
Product Code KRO
UDI-Device Identifier10603295491064
UDI-Public(01)10603295491064(10)M7537M
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K191779
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number151760112
Device Lot NumberM7537M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/10/2025
Supplement Date Received by Manufacturer05/02/2025
Initial Report FDA Received Date05/02/2025
Supplement Report FDA Received Date05/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CONTROL CABLE W/CABLE SLEEVE; LPS DISTAL FEM COMP XXSM LT; LPS FEM TO SLEEVE ADAPTER +5; UNIVERSAL FEM SLV FUL POR 46MM; UNIVERSAL STEM 75X18MM FLUTED
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-