| Catalog Number |
711.13 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported that during surgery ophthalmic forceps tip came off in the patient's eye and they were able to remove it immediately after the incident without any adverse event.The surgery was completed on the same day without any patient harm.Procedure details have not been provided.Additional information has been requested, none has been received till date.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in sections h.6 and h.11.The concerned instrument was not returned to the investigation site for evaluation.Therefore, no further assessment is possible and the investigation is concluded without identifying the root cause.A review of the device history record traceable to the reported lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A non-conformance based review of the lot number was performed and did not reveal any potential contributing factors to the reported complaint.The associated product was not received at the investigation site and not enough information was provided for further investigation.Therefore, the root cause for the customer's reported event cannot be determined conclusively.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.The complaint data will continue to be monitored for evidence of adverse trending and further action will be taken, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|