| Model Number |
AU00T0 |
| Medical Device Problem Code |
Device Difficult to Setup or Prepare (1487)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(6).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that, during intraocular lens implantation surgery, the trailing haptic folded wrong, wrong loaded in shooter, feet first.Hence the lens was taken from the injector and put in another shooter and was implanted.Additional information has been requested and received stating in the surgeon opinion, incorrect position of the lens in the cartridge was cause of the event.There was no impairment to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction information was provided in b.5.Additional information was provided in h.11 the product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.A qualified viscoelastic was indicated.The product investigation could not identify a root cause.The account indicated the product was not available to evaluate, the lens was removed from the device and delivered through an unspecified delivery system.Haptic folding may be influenced by an intermittent plunger rate or if the operating room temperature is too high (greater than 23 degrees centigrade / 73 degrees fahrenheit).If the operating room temperature is too high lens folding consistency is negatively affected as the lens more adherent.This may inhibit lens advancement or contribute to incorrect haptic folding.The instructions for use (ifu) instructs: take at least 7 seconds to gently fold the iol by advancing the plunger forward in one smooth, continuous motion until the front edge of the optic is even with the "fill-to" line on the nozzle.Do not allow any portion of the lens to exit the nozzle tip.At this position the blue spring will contact the clear main body.It is important to not advance the plunger abruptly to fold the iol as improper folding and lens damage may occur.After the lens has been advanced to the "fill-to" line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.No part of the lens should exit the nozzle prior to insertion through the incision.Once the lens is in position at the "fill-to" line, it should be implanted within 3 minutes.Proceeding with implantation of a misfolded haptic or a lens that appears to be "out of position" can result in a broken haptic or other negative outcome.Due diligence has been performed in an attempt to obtain further information on this event.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
There are eight medical device reports associated with this report.This file is 2 of 8.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction information was provided in h.6.On supplementary medical device report (smdr) the fda evaluation codes of b.17 was submitted.It should have been b.20.The manufacturer internal reference number is: (b)(4).
|
| |
|