This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the complaint device was discarded by the customer, therefore not returned to j&j medtech orthopaedics and unavailable for a physical evaluation.A manufacturing record evaluation was performed for the finished device, and no non-conformance was identified.Based on the current available information, we cannot determine a root cause for the reported failure.If additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
|
It was reported that during an unspecified surgical procedure while using the ideal shuttle 45 degrees right -g device the surgeon opened the product and pulled the thread and wire together on the first use and the wire broke.There was no strong resistance, and it broke with a weak force.This does not usually happen.The surgery was completed without any problems using a different product and there was no residual material left inside the body.There was no delay in the procedure reported.There were no reports of injuries, medical intervention or prolonged hospitalization.No additional information was provided.
|